Assessing The Influence Of Electrode Placement On Body Composition Measures From Bioimpedance Technologies

Document Type

Article

Publication Title

Medicine & Science in Sports & Exercise

Abstract

A variety of laboratory-based methods have been utilized to assess body composition. However, technical and physiological variations can influence the estimates from these devices.

PURPOSE: The purpose of this study was to assess the within- and between-day reliability of bioimpedance spectroscopy (BIS) and single-frequency bioelectrical impedance analysis (SFBIA), with varying electrode placement, for estimates of body fat percentage (BF%), fat-free mass (FFM), and fat mass (FM).

METHODS: Eighteen healthy young adults (mean ± SD; age = 25.7 ± 7.5 years; BMI = 22.4 ± 2.3 kg/m2) completed two visits to the laboratory for BIS and SFBIA assessments, separated by 1-2 days. Participants remained supine during tests. For the first visit, duplicate assessments were performed using the same adhesive electrodes without removal (BIS same electrode [BISS], SFBIA same electrode [SFBIAS]). Another assessment was performed after replacing the electrodes (BIS different electrode [BISD], SFBIA different electrode [SFBIAD]). During the second visit, new electrodes were applied, and BIS and SFBIA measures were repeated. Test-retest reliability was established using the absolute technical error of the measurement (TEM) to determine the within-day and between-day errors for BF%, FFM, and FM.

RESULTS: All within-day measures of body composition with the same electrode placement exhibited the lowest absolute TEMs (BF% BISS = 0.06, SFBIAS = 0.03; FFM (kg) BISS = 0.04, SFBIAS = 0.02; FM (kg) BISS = 0.04, SFBIAS = 0.02) followed by within-day different electrode placement (BF% BISD = 0.70, SFBIAD = 0.28; FFM BISD = 0.53, SFBIAD = 0.19; FM BISD = 0.53, SFBIAD = 0.19), and the largest errors were observed for between-day different electrode placement (BF% BISD = 1.24, SFBIAD = 1.07; FFM BISD = 0.90, SFBIAD = 0.85; FM BISD = 0.81, SFBIAD = 0.73).

CONCLUSIONS: Electrode placement is a more important source of within-day technical error for BIS and SFBIA technologies than the analyzers themselves. BIS exhibited greater error due to electrode placement than to between-day biological variability, whereas more of the between-day error in SFBIA estimates was due to biological variability. Both potential technical and biological error must be considered when performing repeated measures using bioimpedance devices.

Research Highlights

  • The Problem: Researchers sought to assess the within-day and between-day reliability of bioimpedance spectroscopy (BIS) and single-frequency bioelectrical impedance analysis (SFBIA) using varying electrode placements to estimate body fat percentage, fat-free mass, and fat mass.

  • The Method: Eighteen healthy young adults with a mean age of 25.7 ± 7.5 years and a mean BMI of 22.4 ± 2.3 kg/m2 completed two laboratory visits separated by one to two days for duplicate supine assessments using identical and replaced adhesive electrodes.

  • Quantitative Finding: Within-day same electrode placement exhibited the lowest absolute technical errors of measurement (BF% BISS = 0.06, SFBIAS = 0.03; FFM (kg) BISS = 0.04, SFBIAS = 0.02; FM (kg) BISS = 0.04, SFBIAS = 0.02); within-day different electrode placement showed higher errors (BF% BISD = 0.70, SFBIAD = 0.28; FFM BISD = 0.53, SFBIAD = 0.19; FM BISD = 0.53, SFBIAD = 0.19); between-day different electrode placement produced the largest errors (BF% BISD = 1.24, SFBIAD = 1.07; FFM BISD = 0.90, SFBIAD = 0.85; FM BISD = 0.81, SFBIAD = 0.73).

  • Finding: Electrode placement constitutes a more significant source of within-day technical error for BIS and SFBIA technologies than the analyzers themselves, with BIS exhibiting greater error from electrode placement and SFBIA reflecting greater error from between-day biological variability.